            LCT LAB-B-B/OIML C3/ NTEP CIII/Bending beam load cell / single-point / For counting scale

            Description

            Model: LAB-B-B

            Brand: LCT

            Capacity: 3-50kg

            Usage: Mainly used for price computing scale,counting scale,weighing scale

            Dimension: 130*22*30mm

            分享

            相關產品

             
             
            Description

            Model: LAB-B-B

            Brand: LCT

            Capacity: 3-50kg

            Usage: Mainly used for price computing scale,counting scale,weighing scale

            Dimension: 130*22*30mm

             



            Features
            • parellel bending beam
            • capacity:3kg,...,50kg
            • accuracy:0.03%R.O.
            • recommended platform size:250*350MM
            • aluminum construction with surface anodized
            • environment protection class:IP65
             
            Applications
            • price computing scales
            • counting scales
            • weighing scales
            • retail scales
            • bench scales
             
            Description
            LCT Single Point Load Cells have been designed such that their excellent mechanical and measurement properties can be optimally used in a wide range of applications. A Single Point Load Cells is also called Platform Load Cell.
            Advantages of LCT Single Point Load Cells / Platform Load Cells:
            Quick to install thanks to off-center load compensation at the factory (per OIML R60),and only one unit is sufficient to build up a scale.
            Model LAB-B-B load cells are designed to this “single point”type and manufactured from high quality aluminum alloy of aviation standard. LAB-B-B load cells can be used to measure loads range from 3kg to 50kg with the measuring accuracy of 0.03% R.O.( R.O.=Rated Output)
            LAB-B-B load cells are mainly used for price computing scales,counting scales,weighing scales,retail scales and bench scale.
             
            Dimension
            Electrical Connection
            Technical Data
             Rated Capacity 3,6,10,15,20,30,40,50 (kg)
            Accuracy Class B *
            Rated Output  2.0±10%mV/V
            Zero Balance ±5%R.O.
            Input Resistance 405±10Ω
            Output Resistance 350±3Ω
            Linearlty Error ±0.02%R.O.
            Repeatability Error ±0.015%R.O.
            Hysteresis Error ±0.015%R.O.
             Creep in 30 Min. ±0.03%R.O.
            Temp.Effect on Output ±0.03%R.O./10℃
            Temp.Effect on Zero ±0.05%R.O./10℃
            Compensated Temp.Range -10-+40℃ 
            Excitation,Recommended 5-12VDC
            Excitation,Maximum  18VDC
            Operating Temp.Range -20-+60℃ 
            Safe Overload 150%R.C.
            Ultimate Overload 200%R.C.
            Insulation Resistance ≥2000MΩ(50VDC)
            Cable,Length ø4mm×0.4m *
            Protection Class IP65